Haemodialysis Machine (HD)
Category: Haemodialysis Equipment
Summary: W-T2008-B haemodialysis machine is designed for performing chronic hemodialysis for adult patients with End-Stage Renal Disease (also known as Chronic Kidney Disease Stage 5).
Product Details
Device Name: Haemodialysis Machine
Class of MDR:IIb
Models:W-T2008-B
Configurations:It is composed of circuit control system,monitoring system,blood extracorporeal circulation control system and hydraulic system,but does not include disposable parts.
Intended use:
The Haemodialysis Machine is designed for performing chronic hemodialysis for adult patients with End-Stage Renal Disease (also known as Chronic Kidney Disease Stage 5).
See the instructions for details
Application Purpose of this Device
W-T2008-B haemodialysis machine is is designed for performing chronic hemodialysis for adult patients with End-Stage Renal Disease (also known as Chronic Kidney Disease Stage 5).
Environment of use:It can be used in limited care centers and clinical hemodialysis.
Forms of Therapy:
Haemodialysis, Isolated Ultrafiltration, Sequential Ultrafiltration, etc.
Features:
Intelligent Double Operation System
LCD touch screen with button interface
Emergency power 30Mins
Blood Pump
Spare Pump (for standby )
Heparin Pump
Hydraulic compartment(Balance chamber + UF pump).
Operation, Alarm information Memory function.
A/B Ceramic proportion pump , High precision, Corrosion-proof, Accuracy.
Introduction
Size & Weight Size: 750mm×620mm×1570mm (L*W*H)
Area: 630*720 mm
Weight: About 88KG
Power Supply 220~240VAC, 50Hz / 60Hz, 16A
Input power: 2000VA
Back-up battery: 30 minutes
Water input pressure: 0.15 MPa ~0.6 MPa
21.75 P.S.I. ~87 P.S.I.
Water input temperature:5℃~35℃
Working environment:Ambient temperature 5ºC~32℃,relative humidty:30%~80%.
Dialysate
Dialysate temperature: preset range 35.0℃~39.0℃
Dialysate flux: 300、500、800 ml/min
Dialysate concentration: 12.8 mS/cm ~15.7 ms/cm, ±0.1 ms/cm
UF rate Flow range: 0 ml/h ~4000 ml/h
Resolution ratio: 1ml
Precision: ±1%
Extracorporeal Part
Venous pressure: -60 mmHg ~+520 mmHg, ±10 mmHg
Arterial pressure : -300 mmHg ~+280 mmHg, ±10 mmHg
TMP pressure : -60 mmHg ~+520 mmHg, ±20 mmHg
Blood pump flow range: 15 ml/min ~400 ml/min(diameter:Ф6 mm)
Spare pump flow range: 15ml/min ~600 ml/min(diameter:Ф8 mm)
Resolution ratio: 1 ml
Precision: error range:± 10%
Heparin Pump
Syringe size: 20, 30, 50 ml
Flow range: 0.1 ml/h ~10 ml/h
Resolution ratio: 0.1ml
Precision: ±5%
Disinfection
Citric acid thermal disinfection:
Time: about 45 minutes
Temperature: ≥85℃
Monitoring System
Dialysate temperature: preset range 35.0℃~39.0℃, ±0.5℃
Blood leak detection: Photochromic
Under the maximum specified dialysate flow rate, ultrafiltration flow rate, the maximum alarm limit of the leakage rate is ≤0.5mL/min (the HCT of blood is 25%), when blood leakage is detected.
Bubble detection: ultrasonic
The alarm will operate when single volume of bubble is bigger than 50 ul at blood flow of 200ml/min. The alarm limit for continuous microbubble passage in the venous circuit is 1 ml/min.
Conductivity: acoustic-optic
Keywords: Haemodialysis Machine (HD)
Previous page
Next page
RECOMMENDED PRODUCTS
Products
PRODUCT FEEDBACK
Product Message